Experimental

 

Measurement Range

5 x 104 - 4 x 106 cells per ml

Analysis Time

15 seconds to 2 minutes

Loading Volume

20µl per test

Measurement Volume

9µl per test

Disposable Chips

Two Channel chip (200 tests / kit)
  Four Channel chip (400 tests / kit)

Staining Solution

Propidium Iodide

Hardware

 

Weight

9KG

Size

220 x 375 x 250 mm

Power

AC100-240V, 50 - 60 Hz

Objective Lens

x 4

Excitation Source

Green laser (532nm)

CCD Camera

B/W CCD

Filter

Emisssion Filter

 