ELx800™ Absorbance Microplate Reader
|
|
Specifications
| Detection method | Absorbance |
| Read method | Endpoint Kinetic and well area scanning via computer control |
| Microplate types | 6-, 12-, 24-, 48-, 96-well plates 384- and 60/72/96-well Terasaki plates (NB option) |
| Reading speed | 30 secs single wavelength |
| Wavelength range | 400 to 750nm 340 to 750nm (UV option) |
| Wavelength selection | Filters |
| Filter wheel capacity | 5 positions |
| Filters supplied | 4 filters (5 with UV option) |
| Bandpass | 10 nm |
| Dynamic range | 0 to 3.0 OD |
| OD accuracy | <1% at 2.0 OD typical |
| OD linearity | <1% at 2.0 OD typical <3% at 3.0 OD typical |
| OD repeatability | <0.5% at 2.0 OD typical |
| Resolution | 0.001 OD |
| Light source | Tungsten Halogen |
| Onboard software | 55 user-programmable protocols |
| PC Software | KCjunior, Gen5 (optional) |
| Power | 100-240 VAC 50/60 Hz |
| Dimensions | 16.5" D x 15" W x 7" H (41.9 x 38.1 x 17.8 cm) |
| Weight | 18.5 lbs (8 kg) |
| Regulatory | IVD (In Vitro Diagnostics) Directive. All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek. |