PowerWave™ Microplate Spectrophotometer
BioTek

PowerWave™ Microplate Spectrophotometer PowerWave™ Microplate Spectrophotometer
bullet off Overview
bullet off Features
bullet off Models
bullet off Specifications
bullet off Accessories
bullet off Print Product Details

bullet on Specifications


Detection method Absorbance
Read method Endpoint, kinetic, spectral, linear and well area scanning (depending on model)
Microplate types 96- and 384-well microplates (PowerWave and PowerWave HT models);
6- to 384-well microplates and 60-/72-/96-well Terasaki plates (PowerWave XS model);
Also reads up to 8 Bio-Cells for 1 cm measurement.
Reading speed PowerWave and PowerWave HT:
  • Normal: 96 = 20 sec, 384 = 66 sec
  • Rapid: 96 = 11 sec, 384 = 23 sec
  • Sweep: 96 = 5 sec, 384 = 11 sec

    PowerWave XS:
  • Normal: 96 = 49sec, 384 = 150 sec
  • Rapid: 96 = 40 sec, 384 = 112 sec
  • Sweep: 96 = 14 sec, 384 = 26 sec
  • Wavelength range 200 to 999 nm;
    340 to 999 nm (PowerWave HT 340 and PowerWave 340 models)
    Wavelength selection Monochromator, selectable 1 nm or greater increments
    Monochromator wavelength accuracy +2 nm
    Monochromator wavelength repeatability +0.2 nm
    Bandpass 5 nm (PowerWave and PowerWave HT)
    2.4 nm (PowerWave XS)
    Dynamic range 0 to 4.0 OD
    OD accuracy 1% + 0.01 OD typical
    OD linearity +1% typical
    OD repeatability 0.5% + 0.005 OD typical
    Resolution 0.001 OD
    Light source Xenon flash lamp
    Stray light 0.03% at 230 nm typical
    Temperature control Ambient +4oC to 50oC
    4-Zone™ natural convection
    Shaking User-programmable timing and frequency
    PC Software Includes Gen5 or KCjunior software (depending on model)
    Power 100 VA max;
    Instrument powered by universal input (100-240 VAC 50-60 Hz) power supply.
    Dimensions 16" D x 8.5" W x 8.5" H 
    (40.6 x 21.6 x 21.6 cm)
    Weight 24 lbs (11 kg)
    Regulatory For in-vitro diagnostic use (Except in the European Union).
    All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek.
    Automation Compatible with Bio-Stack Microplate Stacker and Bio-Stack Twister II Microplate Handler.

    Specifications subject to change.

     

    Suppliers | Site Map | Products | Contact Us | ©2006 Labtech International LTD | Telephone 01273 814 888 | Email sales@labtech.co.uk