 General
Detection methodFluorescence, Time Resolved Fluorescence, Fluorescence Polarization, Luminescence, UV-Visible Absorbance
Read methodEnd-point, kinetic, spectral scanning, well-area scanning
Microplate typesMonochromator system 1- to 384-well plates
Filter system 1- to 1536-well plates (luminescence 1-384)
PCR plates
Temperature control4ºC above ambient to 50ºC ± 0.5ºC at 37ºC
ShakingYes
AutomationYes
SoftwareGen5T Included
Control through USB or serial port
 Absorbance
Light sourceSQ Xenon Flash
Wavelength selectionMonochromator
Wavelength range200 - 999 nm, 1 nm increment
Bandpass2.4 nm
Dynamic range0 - 4.0 OD
Resolution0.0001 OD
Pathlength correctionYes
Monochromator wavelength accuracy+/- 2 nm
Monochromator wavelength repeatability+/- 0.2 nm
OD accuracy< 1% at 2.0 OD typical
< 3% at 3.0 OD typical
OD linearity< 1% from 0 to 3.0 OD typical
OD repeatability< 0.5% at 2.0 OD typical
Stray light0.03% at 230 nm typical
Reading speed96: 11 seconds
384: 22 seconds
1536: 42 seconds
 Fluorescence Intensity
Light sourceTungsten Halogen
High Energy DPR Xenon Flash
SensitivityMonochromators: fluorescein 5 pM typical (0.5 fmol/well 384-well plate)

Filters/mirrors: fluorescein 1 pM typical (0.1 fmol/well 384-well plate)
Wavelength selectionDouble grating monochromators (Top) and,
Deep blocking bandpass filters / dichroic mirrors (Top/Bottom)
Wavelength rangeMonochromators: 250 - 800 nm
Filters: 200 - 700 nm (900 nm option)
Dynamic range> 6 decades
Reading speed96: 11 seconds
384: 22 seconds
1536: 43 seconds
 Luminescence
Sensitivity10 amol ATP typical (flash)
Wavelength range300 - 700 nm
Dynamic range> 6 decades
 Fluorescence Polarization
Sensitivity3 mP at 1 nM fluorescein typical
Wavelength selectionDeep blocking bandpass filters / dichroic mirrors (Top)
Wavelength range300 - 700 nm (900 nm option)
 Time-Resolved Fluorescence
Light sourceHigh Energy DPR Xenon Flash
SensitivityEuropium 60 fM typical with filters (6 amol/well in 384-well plate)
Wavelength selectionDouble grating monochromator (Top) and,
Deep blocking bandpass filters / dichroic mirrors (Top / Bottom)
Wavelength rangeFilters: 200 - 700 nm (900 nm option)
Monochromators: 250 - 850 nm
 Reagent Dispensers
Number2 syringe pumps
Dispense volume5 - 1000 µl in 1 µl increment
Dead volume1.1 ml, 100 µl with backflush
 Physical Characteristics
Power250 Watts max.
Dimensions17"W x 20.9"D x 15"H
(43.5 x 53.1 x 38.1 cm)
Weight78 lbs (35 kg)
 Regulatory
RegulatoryFor in vitro diagnostic use (Except in the European Union).
All BioTek microplate instrumentation is CE and NRTL Safety Agency marked. For details on approvals and standards compliance, please contact BioTek.