NanoDrop ND-8000 Eight Sample UV/Vis 1µl Spectrophotometer
|
Sample Type |
Detection Limit (pmol/ul) |
Approximate Upper Limit (pmol/ul) |
Typical Reproducibility (minimum 5 replicates) (SD= pmol/ul; CV= %) |
Cy3, Cy3.5, Alexa Fluor 555, Alexa Fluor 660 |
0.20 |
100 |
sample range 0.20-4.0 pmol/ul: ± 0.20 pmol/ul sample range >4.0 pmol/ul: ± 2% |
Cy5, Cy5.5 and Alexa Fluor 647 |
0.12 |
60 |
sample range 0.12-2.4 pmol/ul: ± 0.12 pmol/ul sample range >2.4 pmol/ul: ± 2% |
Alexa Fluor 488 and Alexa Fluor 594 |
0.40 |
215 |
sample range 0.40-8.0 pmol/ul: ± 0.40 pmol/ul sample range >8.0 pmol/ul: ± 2% |
Alexa Fluor 546 |
0.30 |
145 |
sample range 0.30-6.0 pmol/ul: ± 0.30 pmol/ul sample range >6.0 pmol/ul: ± 2% |