NanoDrop ND1000

 E-mail this product to a friend


     

Overview

The NanoDrop® ND-1000 UV-Vis Spectrophotometer enables highly accurate analyses of extremely small samples with remarkable reproducibility. The patented sample retention system eliminates the need for cuvettes and capillaries which decreases the measurement cycle time. In addition, the high absorbance capability eliminates the need for most dilutions. You’ll find yourself making fewer dilutions and taking more readings than ever before, essentially becoming a more quantitative laboratory.
 
The ND-1000 offers many benefits over other spectrophotometers:
  • Small samples: designed for 1 ul samples
  • High absorbance capability: 50 times that of traditional spectrophotometers
  • No Dilutions for most samples
  • Full Spectrum (220-750nm)
  • 10 second measurement time
  • No cuvettes or capillaries
  • Small size (size of a tissue box)
  • Easy set up (plug and play)
bullet on Specifications
  • 1mm path length (with auto-ranging to 0.2mm)
  • 1 microliter sample
  • Xenon flash lamp
  • 2048-element linear silicon CCD array detector
  • Wavelength range: 220-750 nm
  • Wavelength accuracy: 1nm
  • Wavelength resolution: 3 nm (FWHM at Hg 546 nm)
  • Absorbance precision: .003 absorbance (1mm path)
  • Absorbance accuracy: 2% (at .76 absorbance at 257 nm)
  • Absorbance range: 0.02-75 (10mm equivalent absorbance)
  • Detects nucleic acid down to 2 ng/microliter
  • Measures nucleic acid up to 3700 ng/microliter (without dilution)
  • Measurement cycle: 10 seconds
  • Dimensions: 15 cm x 20 cm x 12 cm
  • Weight: 3 kg
  • Operating voltage: 12 vdc
  • Operating power consumption: 6 W
  • Standby power consumption: 1.5 W
  • CE approval (units sold in Europe, Australia and New Zealand)

    Usermanual




  •