PA-4000/PA-201BS Programmable Pulse Switch System

 E-mail this product to a friend


     

PA-4000/PA-201BS Programmable Pulse Switch System (Computer Required)
View Data Sheet (66 KB)

PA 201The PA-201 Programmable Pulse Switch option is used in conjunction with the PA-4000 electroporator to expand the available number of electrode ports from one as found in a standard cuvette to eight total. Opening this window enables researchers to explore new and exciting electrical field delivery techniques in the search for an optimal process. In particular, reverse polarity cuvette electroporation and multi-probe in-vivo electroporation are two techniques that are the subject of upcoming breakthroughs.

This system provides the greatest control and process feedback available, while retaining the ability to modularize further into applications such as cell fusion or high throughput electroporation. In addition, the intuitive Pulse Agile Interface software quickly guides even first-time users to direct success. Superior results are consistently achieved as a result of our stringent structural requirements and testing process.

PA 4000-201

  • Precision rectangular (square) wave electroporation
  • Dynamically programmable 8-port output
  • Banana plug output to cuvette chamber for reverse polarity (CE-24 Only)
  • Computer controlled operation
  • PulseAgile® protocol creation
  • Log file automatically saved after every run
  • Ionic solutions down to 10 ohms (PBS, etc.)
  • Automatic safety shutdown on arc detection
  • Built-in voltage monitor for oscilloscope verification

Click here for Larger Image

There are a number of patents on applications that may use this type of system and it is the responsibility of the user not to infringe on those patents.

PA 210 graph8-Port Output

The PA-201 is an eight-port array output switch, of which each port can be dynamically programmed via software to become an anode, cathode, or left unconnected during run-time. Various configurations have been found to accommodate a more efficient delivery when compared to two-point electrodes, particularly for in-vivo tumor applications.

Technical Reference: Sersa, G. et al, Bioelectrochem. Bioenerg. 39, 61-66 (1996)

Reverse Polarity

A new technology yet to be fully unlocked, reversing polarity of electric fields between pulses theoretically indicates a significant improvement over standard electroporation. This relies upon promoting molecule movement in multiple directions as well as evening out the distribution of membrane poration around cells (single-direction fields traditionally open larger pores on the anode side).

Rectangular (square) Wave

Rectangular wave electroporators have become industry standard because of their ability to increase transfection efficiency while maintaining viability over older and less reliable exponential decay models. Users are given the full ability to adjust parameters for voltage (electric field), pulse width, pulse number, and interval between pulses in an easy to navigate menu system.

PulseAgile® Protocol

The Cyto Pulse Sciences patented PulseAgile® protocol series takes the concept of square wave electroporation one step further: users are given the choice to create multiple protocols with different parameters, called groups, that are run in direct succession of one another. This is generally used to deliver a series of high voltage and short duration pulses first permeabilizing the cell membrane, followed by a series of low voltage and long duration pulses to continue delivery of the transfectant by electrophoresis.

The ultimate goal of creating these types of protocols is to deliver similar transfection efficiencies while retaining an even higher viability than standard four-parameter waves. In particular, published in-vivo applications have repeatedly been found to substantially improve delivery in addition to some ex-vivo applications.

System Monitoring

There is no mystery to what’s inside the box when all necessary information can be accessed from the outside included as standard features is a voltage monitors ready to attach with an external oscilloscope in order to view real-time pulsing conditions. Using this connection is a true safeguard to know exactly how the applied waveform reflects the programmed waveform, so you never again have to wonder if your protocol ran as intended.

Fault Detection

All Cyto Pulse Sciences equipment is designed to withstand extreme user conditions. In the event of a fault, advanced control circuitry immediately detects the source of any problem and proceeds to halt the system to prevent permanent damage. The safety of Cyto Pulse customers is our number one priority.

  • Electrode Interlock - improperly attached cuvettes or electrodes will cause the protocol start function to be disabled until corrected
  • Internal Fault - damaged internal system components will cause high voltage to be turned off and fault screen to be displayed

Cuvette Holder

The heavy-duty electroporation cuvette holder will operate with all major cuvette manufacturers, although Cyto Pulse cuvettes are recommended for best results; standard cuvette sizes are available in 1 mm, 2 mm, and 4 mm electrode gaps. The insertion mechanism is notched to prevent incorrect positioning and only upon full insertion will high voltage become enabled to pulse. A large plastic barrier surrounds the cuvette field to completely block any contact with the electrodes.  The CE-20 is standard, the CE-24 is required for polarity reversal

 

Specifications

General  
      Pulse Waveform Rectangular (Square)
      Interface Computer with Windows®
      Electrode Cuvette, In-vivo
Protocol  
      Pulse Amplitude HV Mode 20 - 1100 volts @ 5 volt step
      Pulse Amplitude LV Mode 20 - 400 volts @ 2 volt step
      Pulse Amplitude Accuracy ±5%
      Max E-Field (1 mm cuvette) 11 kV/cm
      Max E-Field (2 mm cuvette) 5.5 kV/cm
      Max E-Field (4 mm cuvette) 2.75 kV/cm
      Pulse Duration 0.001 E100 ms (<300 volts)
0.001 E40 ms (<400 volts)
0.001 E20 ms (<500 volts)
0.001 E1 ms (<1100 volts)
      Pulse Interval 0.125 - 400 sec
      Number of Pulses 1 - 99
      Number of Pulse Groups 1 - 40
      Polarity Reversal Yes
Safety / Monitoring  
      Electrode Interlock Yes
      Pulse Voltage Monitor BNC Output
Electrical  
      AC Line 100 - 240 VAC 50/60Hz
Physical  
      High Voltage Outputs MHV, Banana Plug
      Computer Control RS232
      Dimensions (W-D-H)

PA-4000
12.6" x 11.8" x 8.2"
(32 cm x 30 cm x 20.8 cm)

PA-201
12.6" x 11.8" x 3"
(32 cm x 30 cm x 7.5 cm)

      Weight

PA-4000
23 pounds (11 kg)

PA-201
6.5 pounds (3 kg)

 

Included in the system
PA-4000/PA-201S
PA-4000
PulseAgile® Advanced Rectangular Wave Electroporator
PA4-SW
PulseAgile® Application Software
PA4-UMAN
PA-4000 User Manual
CE-20
Cuvette Holder
CUV-M

Cuvette Multi-pack set of 5 each

1 mm, 2 mm, 4 mm

CS-L-XX
IEC Line Cord, Serial cable, XX=country
PA-201
Programmable Pulse Switch
PA-201-UMAN
PA-201 User Manual
CS-OPT
Interface Cable Set: DB25 cable, HV cable
PA-201-C
Output cable 9-wire
 